share:

This chromatogram may contain analytes that are synonyms of your search terms.

GC Potency Analysis for Marijuana on Rxi-5Sil MS

PeakstR (min)Conc.
(wt.%)
1.CBD4.0350.0
2.THC4.5243.6
3.CBN4.8400.3
image
GC_GN1156
ColumnRxi-5Sil MS, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13620)
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:Premium 4.0 mm ID single taper inlet liner w/wool (cat.# 23303.5)
Inj. Temp.:250 °C
Oven
Oven Temp.:200 °C (hold 0 min) to 300 °C at 15 °C/min (hold 0 min)
Carrier GasH2, constant pressure (7 psi, 48.3 kPa)
Temp.:200 °C
Dead Time:0.6 min @ 200 °C
DetectorFID @ 300 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
InstrumentHP5890 GC
NotesBlue trace = cannabinoids standard (cat.# 34014) diluted to 100 µg/mL in isopropyl alcohol.
Red trace = extracted marijuana sample

Sample extraction: Weigh 0.2 g of sample into a 40 mL VOA vial, add 40 mL of isopropyl alcohol, shake for 5 minutes, and allow sample to settle.

Quantification: Potency values (weight%) were based on a 1-point standard curve using the standard show above.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

marijuana

,

Rxi-5SilMS

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us