share:

Custom Cleaning Solvents Standard on Stabilwax-MS

Peaks
1.2-Methoxy-ethanol
2.BFB (IS)
3.2-Butoxy ethanol
Peaks
4.2-Butoxyethyl acetate
5.Diethylene glycol ethyl ether acetate
6.Diethylene glycol butyl ether
image
GC_GN1164
ColumnStabilwax-MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 10673)
SampleCustom standard
Diluent:Methanol
Conc.:200 ppm
Injection
Inj. Vol.:1.0 µL split (split ratio 300:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.5)
Inj. Temp.:200 °C
Oven
Oven Temp.:60 °C to 220 °C at 30 °C/min
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
12.330-200
Transfer Line Temp.:200 °C
Analyzer Type:Quadrupole
Source Temp.:200 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Solvent Delay Time:2.3 min
Tune Type:PFTBA
InstrumentAgilent 7890A GC & 5975C MSD
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Stabilwax

,

glycols

,

,

wax

,

cleaning solvents

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us