share:

Residual Solvents on Rxi-624Sil MS via HS-Syringe-GC-HES-MS

PeakstR (min)Conc.
(µg/mL)
1.Propane1.1410
2.Isobutane1.3010
3.n-Butane1.4510
4.Methanol1.6910
5.n-Pentane2.3010
6.Ethanol2.4510
7.Acetone2.9110
8.Isopropanol3.1910
9.Acetonitrile3.3110
10.n-Hexane4.2310
11.Tetrahydrofuran5.2910
12.Chloroform5.6910
13.Carbon tetrachloride5.7010
14.Benzene5.9110
15.n-Heptane6.2510
16.Toluene8.0310
17.m-Xylene10.0610
18.p-Xylene10.0610
19.o-Xylene10.3910
20.Solvent (N,N-dimethylacetamide [DMA])10.64
image
GC_GN1197
ColumnRxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868)
Sample
Diluent:RO water
Conc.:10 µg/mL
Injection
Inj. Vol.:500 µL headspace-syringe split (split ratio 100:1)
Liner:Topaz 0.75 mm ID straight/SPME inlet liner (cat.# 23434)
Inj. Temp.:280 °C
Headspace-Syringe
Split Vent Flow Rate:3 mL/min
Instrument:PAL RTC
Syringe Temp.:135 °C
Sample Temp.:85 °C
Sample Equil. Time:2 min
Inj. Speed:500 µL/sec
Oven
Oven Temp.:30 °C (hold 3 min) to 85 °C at 15 °C/min (hold 2 min) to 250 °C at 35 °C/min (hold 2 min)
Carrier GasHe, constant flow
Flow Rate:2.0 mL/min
Linear Velocity:51 cm/sec
DetectorHES-MS
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1029-4506.4
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Type:HES
Aperture Size:6 mm ID
Source Temp.:350 °C
Quad Temp.:200 °C
Electron Energy:70 eV
Solvent Delay Time:0 min
Tune Type:PFTBA
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977B MSD
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Rxi-624Sil

,

Residual

,

Solvents

,

Cannabis

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us