share:

This chromatogram may contain analytes that are synonyms of your search terms.

Residual Solvents #1 on Rxi-624Sil MS

PeakstR (min)Conc.
(µg/mL)
1.Methanol2.12510
2.n-Pentane3.01810
3.Ethanol3.21210
4.Diethyl ether3.34410
5.Acetone3.82010
6.2-Propanol4.18810
7.Acetonitrile4.36710
8.Methylene chloride4.71010
9.n-Hexane6.03310
10.Ethyl acetate7.67510
11.Chloroform8.10410
12.Cyclohexane8.32810
13.Benzene8.76310
14.1,2-Dichloroethane8.87210
15.n-Heptane9.23610
16.Trichloroethene9.63710
17.Toluene11.50010
18.m-Xylene13.42110
19.p-Xylene13.42110
20.o-Xylene13.72510
21.N,N-Dimethylacetamide (solvent)13.958
image
GC_GN1199
ColumnRxi-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868)
SampleResidual solvents #1 (cat.# 34105)
Diluent:N,N-Dimethylacetamide
Conc.:10 µg/mL
Injection
Inj. Vol.:250 µL headspace-syringe split (split ratio 10:1)
Liner:Topaz 1.8 mm ID straight/SPME inlet liner (cat.# 23280)
Inj. Temp.:280 °C
Headspace-Syringe
Split Vent Flow Rate:20 mL/min
Instrument:PAL RTC
Syringe Temp.:150 °C
Sample Temp.:80 °C
Sample Equil. Time:45 min
Inj. Speed:1700 µL/sec
Oven
Oven Temp.:30 °C (hold 6 min) to 85 °C at 15 °C/min (hold 2 min) to 250 °C at 35 °C/min
Carrier GasHe, constant flow
Flow Rate:2 mL/min
Linear Velocity:40.63 cm/sec
DetectorFlame Ionization Detector @ 320 °C
Make-up Gas Flow Rate:40 mL/min
Make-up Gas Type:N2
Hydrogen flow:45 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent 7890B GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Cannabis

,

Residual

,

Solvents

,

Analysis

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us