share:

This chromatogram may contain analytes that are synonyms of your search terms.

Helium/Hydrogen Separation using Rt-Msieve 5A

Peaks
1.Helium
2.Hydrogen
image
GC_MS01136
ColumnRt-Msieve 5A, 30 m, 0.53 mm ID, 50 µm (cat.# 19723)
SampleHelium/hydrogen gas mixture
Injection
Inj. Vol.:25 µL split (split ratio 10:1)
Liner:2 mm split Precision liner w/wool (cat.# 20823)
Inj. Temp.:200 °C
Split Vent Flow Rate:40 mL/min
Oven
Oven Temp.:30 °C
Carrier GasN2, constant pressure (7.0 psi, 48.3 kPa)
Linear Velocity:38 cm/sec @ 30 °C
DetectorTCD @ 200 °C
Make-up Gas Type:N2
Data Rate:20 Hz
Sensitivity Mode:N2/Ar
InstrumentHP5890 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Helium

,

Hydrogen

,

Rt-Molsieve 5A

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us