share:

This chromatogram may contain analytes that are synonyms of your search terms.

Sim Dist, ASTM D3710 (Gasoline) on Rtx-1 SimDist 2887

Peaks
1.n-Propane
2.2-Methyl propane
3.n-Butane
4.2-Methylbutane
5.n-Pentane
6.2-Methylpentane
7.n-Hexane
8.2,4-Dimethylpentane
9.n-Heptane
10.Toluene
11.p-Xylene
12.n-Propylbenzene
13.n-Decane
14.n-Butylbenzene
image
GC_PC00230
ColumnRtx-1 SimDist 2887, SilcoSmooth Tubing, 25 inches, 1/8 in. OD, 2 mm ID (cat.# 80000-800)
Sampleunleaded gasoline (ASTM D2887)
Injection
Inj. Vol.:1.0 µL packed not on-column
Inj. Temp.:250 °C
Oven
Oven Temp.:-30 °C to 250 °C at 10 °C/min
Carrier GasHe, constant flow
Flow Rate:25 mL/min
DetectorFID @ 300 °C
NotesFID sensitivity: 256 x 10-11 AFS
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

packed column

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us