Crude Oil, Prudhoe Bay on Rtx-1
Peaks | |
---|---|
1. | Benzene![]() |
2. | Heptane![]() |
3. | Toluene![]() |
4. | Octane![]() |
5. | m-Xylene/p-Xylene![]() |
6. | o-Xylene![]() |
7. | Nonane![]() |
8. | Decane![]() |
9. | Undecane![]() |
10. | Dodecane![]() |
11. | Tridecane![]() |
12. | Tetradecane![]() |
13. | Pentadecane![]() |
14. | Hexadecane![]() |
Peaks | |
---|---|
15. | Heptadecane![]() |
16. | Pristane![]() |
17. | Octadecane![]() |
18. | Phytane![]() |
19. | Nonadecane![]() |
20. | Eicosane![]() |
21. | Heneicosane![]() |
22. | Docosane![]() |
23. | Tricosane![]() |
24. | Tetracosane![]() |
25. | Pentacosane![]() |
26. | Hexacosane![]() |
27. | Heptacosane![]() |
28. | Octacosane![]() |

GC_PC00231
Column | Rtx-1, 30 m, 0.53 mm ID, 3.00 µm (cat.# 10185) |
Sample | Prudhoe Bay crude oil |
Injection | |
Inj. Vol.: | 0.4 µL split (split ratio 50:1) |
Inj. Temp.: | 285 °C |
Oven | |
Oven Temp.: | 40 °C to 285 °C at 5 °C/min (hold 30 min) |
Carrier Gas | H2, constant flow |
Linear Velocity: | 40 cm/sec |
Detector | FID @ 285 °C |
Notes | FID sensitivity: 256 x 10-11 AFS Flow rate: 5.2 cc/min |
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.