Permanent Gases and C1+C2 Hydrocarbons on ShinCarbon ST (micropacked)
Peaks | |
---|---|
1. | Oxygen![]() |
2. | Nitrogen![]() |
3. | Carbon monoxide![]() |
4. | Methane![]() |
5. | Carbon dioxide![]() |
6. | Acetylene![]() |
7. | Ethylene![]() |
8. | Ethane![]() |

GC_PC00665
Column | 100/120 mesh, ShinCarbon ST, 2 m, 1/16 in. OD, 1.0 mm ID (cat.# 19808) |
Sample | Permanent gases-C1+C2 hydrocarbons |
Injection | |
Inj. Vol.: | 5 µL packed on-column |
Inj. Temp.: | 100 °C |
Oven | |
Oven Temp.: | 40 °C (hold 3 min) to 250 °C at 8 °C/min (hold 10 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 10 mL/min |
Detector | HID @ 200 °C |