share:

This chromatogram may contain analytes that are synonyms of your search terms.

Refinery Gas #6 on Rt-QS-BOND

Peaks
1.Methane
2.Ethane
3.Propylene
4.Propane
5.Isobutane
6.1-Butene
Peaks
7.Isobutylene
8.n-butane & cis-2-butene
9.trans-2-Butene
10.Isopentene
11.1-Pentene
12.n-pentane & cis-2-pentene
13.n-Hexane
image
GC_PC01014
ColumnRt-QS-BOND, 30 m, 0.53 mm ID, 20 µm (cat.# 19738)
Samplerefinery gas mix (mole%)
Injection
Inj. Vol.:20 µL split (split ratio 10:1)
Liner:Siltek Gooseneck Splitless (4mm)
Inj. Temp.:240 °C
Oven
Oven Temp.:40 °C (hold 2 min) to 225 °C at 15 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:5.7 mL/min
DetectorFID @ 240 °C
NotesSiltek Gooseneck Splitless (4mm) Obsolete, recommend catalog number 20709
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

refinery gas #6

,

refinery gas

,

hydrocarbon gases

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us