share:

This chromatogram may contain analytes that are synonyms of your search terms.

Butane Lighter Fluid on Rt-Alumina BOND/Na2SO4

Peaks
1.Methane
2.Ethane
3.Propane
4.Isobutane
5.n-Butane
6.trans-2-Butene
7.1-Butene
8.Isobutylene
image
GC_PC01083
ColumnRt-Alumina BOND/Na2SO4, 50 m, 0.53 mm ID, 10 µm (cat.# 19756)
SampleButane lighter fluid
Injection
Inj. Vol.:100 µL split
Liner:Splitless taper (2 mm) (cat.# 20795)
Inj. Temp.:200 °C
Split Vent Flow Rate:400 mL/min
Oven
Oven Temp.:45 °C (hold 1 min) to 200 °C at 10 °C/min (hold 3.5 min)
Carrier GasH2, constant pressure (8.0 psi, 55.2 kPa)
Linear Velocity:74 cm/sec @ 45 °C
DetectorFID @ 200 °C
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentHP5890 GC
Notes100 μL sample loop on gas sampling valve, split
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Butane

,

Lighter Fluid

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us