Butane Lighter Fluid Rt-Alumina BOND/KCl

Peaks
1.Methane
2.Ethane
3.Propane
4.Isobutane
5.n-Butane
6.trans-2-Butene
7.1-Butene
8.Isobutylene
image
GC_PC01086
ColumnRt-Alumina BOND / KCl, 50 m, 0.53 mm ID, 10 µm (cat.# 19760)
Samplebutane lighter fluid
Injectionsample valve
Sample Loop Vol.:100 µL
Valve Name:sample loop on gas sampling valve, split (split vent flow 400mL/min.), 2mm single gooseneck liner (cat.# 20795)
Valve Temp.:200 °C
Oven
Oven Temp.:45 °C (hold 1 min) to 200 °C at 10 °C/min (hold 3.5 min)
Carrier GasH2, constant pressure (8.0 psi, 55.2 kPa)
Linear Velocity:74 cm/sec @ 45 °C
DetectorFID @ 200 °C