share:

Propane Camping Fuel on Rt-Alumina BOND/KCl

Peaks
1.Methane
2.Ethane
3.Ethylene
4.Propane
5.Propylene
6.Isobutane
7.n-Butane
image
GC_PC01087
ColumnRt-Alumina BOND/KCl, 50 m, 0.53 mm ID, 10 µm (cat.# 19760)
Samplepropane camping fuel
Injection
Inj. Vol.:100 µL split
Liner:Gooseneck Splitless (2mm) (cat.# 20795)
Inj. Temp.:200 °C
Split Vent Flow Rate:400 mL/min
Oven
Oven Temp.:45 °C (hold 1 min) to 200 °C at 10 °C/min (hold 3.5 min)
Carrier GasH2, constant pressure (8..0 psi, 55.2 kPa)
Linear Velocity:74 cm/sec @ 45 °C
DetectorFID @ 200 °C
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentHP5890 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Rt®-Alumina BOND/KCl

,

propane

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.