Natural Gas on Rt-Alumina BOND/Na2SO4

Peaks
1.Methane
2.Ethane
3.Propane
4.Isobutane
5.n-Butane
6.neo-Pentane
7.iso-Pentane
8.n-Pentane
9.Hexane isomers
10.Heptane isomers
11.Benzene
12.Octane isomers
image
GC_PC01092
ColumnRt-Alumina BOND/Na2SO4, 50 m, 0.53 mm ID, 10 µm (cat.# 19756)
SampleNatural gas
Injection
Inj. Vol.:500 µL split
Liner:2.0 mm ID single taper (cat.# 20795)
Inj. Temp.:200 °C
Split Vent Flow Rate:50 mL/min
Oven
Oven Temp.:45 °C (hold 1 min) to 200 °C at 10 °C/min (hold 8.5 min)
Carrier GasH2, constant pressure (8.0 psi, 55.2 kPa)
Linear Velocity:45 cm/sec @ 45 °C
DetectorFID @ 200 °C
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentHP5890 GC