share:

ASTM Method D3606 on D3606 Application Column Set

PeaksConc.
(vol.%)
1.C761%
2.Ethanol10%
3.Benzene5%
4.2-Butanol (IS)4%
5.Toluene20%
image
GC_PC01103
ColumnD3606 Application Column, 2 column set
Column 1: nonpolar Rtx-1, 6' (1.8m), 1/8" OD, 2.0mm ID
Column 2: proprietary packing material, 16' (4.9m), 1/8" OD, 2.0mm ID (cat.# 83606-800)
Samplemixed standard
Diluent:C7
Injection
Inj. Vol.:1.0 µL Direct
Inj. Temp.:200 °C
Oven
Oven Temp.:135 °C
Carrier GasHe, constant flow
Flow Rate:25 mL/min
DetectorTCD @ 200 °C
InstrumentWasson 3606/5599 combo 6890N GC
NotesBackflush time: 3 min.

RELATED CHROMATOGRAM LIBRARY SEARCHES

ASTM D3606

,

ethanol

,

wasson

,

ASTM

,

D3606

,

benzene

,

packed column

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us