ASTM Method D3606 on D3606 Application Column Set

PeaksConc.
(vol.%)
1.C761%
2.Ethanol10%
3.Benzene5%
4.2-Butanol (IS)4%
5.Toluene20%
image
GC_PC01103
ColumnD3606 Application Column, 2 column set
Column 1: nonpolar Rtx-1, 6' (1.8m), 1/8" OD, 2.0mm ID
Column 2: proprietary packing material, 16' (4.9m), 1/8" OD, 2.0mm ID (cat.# 83606-800)
Samplemixed standard
Diluent:C7
Injection
Inj. Vol.:1.0 µL Direct
Inj. Temp.:200 °C
Oven
Oven Temp.:135 °C
Carrier GasHe, constant flow
Flow Rate:25 mL/min
DetectorTCD @ 200 °C
InstrumentWasson 3606/5599 combo 6890N GC
NotesBackflush time: 3 min.