share:

Ethylene and C1-C5 Hydrocarbons by ASTM D6159-97 on Rt-Alumina BOND/KCl, Rtx-1

Peaks
1.Methane
2.Ethane
3.Ethylene
4.Propane
5.Cyclopropane
6.Propylene
7.Acetylene
8.Isobutane
9.Propadiene
Peaks
10.n-Butane
11.trans-2-Butene
12.1-Butene
13.Isobutylene
14.cis-2-Butene
15.iso-Pentane
16.Methylacetylene
17.n-Pentane
18.1,3-Butadiene
image
GC_PC01110
ColumnRt-Alumina BOND/KCl *, 50 m, 0.53 mm ID, 10 µm (cat.# 19760)
SampleEthylene gas plus C1 through C5 hydrocarbons
Injection
Inj. Vol.:1 µL split
Liner:2 mm splitless (cat.# 20712)
Inj. Temp.:200 °C
Split Vent Flow Rate:60 ml/min
Oven
Oven Temp.:35 °C (hold 2 min) to 190 °C at 4 °C/min (hold 15 min)
Carrier GasHe, constant pressure (8.0 psi, 55.2 kPa)
Linear Velocity:25.4 cm/sec @ 35 °C
DetectorFID @ 200 °C
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentHP5890 GC
Notes* Rt-Alumina BOND/KCl, 50 m, 0.53 mm ID, 10.0 μm (cat.# 19760) in series with an Rtx-1, 30 m, 0.53 mm ID, 5.0 μm (cat.# 10179) connected using a universal Press-Tight connector (cat. # 20401)

(conditions as per ASTM D6159-97)
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Rt-Alumina BOND/KCL

,

Ethylene

,

ASTM D6159-97

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us