Propylene and C1-C5 Hydrocarbons on Rt-Alumina BOND/Na2SO4

Peaks
1.Methane
2.Ethane
3.Ethylene
4.Propane
5.Cyclopropane
6.Propylene
7.Isobutane
8.n-Butane
9.Propadiene
Peaks
10.Acetylene
11.trans-2-Butene
12.1-Butene
13.Isobutylene
14.cis-2-Butene
15.Isopentane
16.n-Pentane
17.1,3-Butadiene
18.Methylacetylene
image
GC_PC01142
ColumnRt-Alumina BOND/Na2SO4, 50 m, 0.53 mm ID, 10 µm (cat.# 19756)
SamplePropylene and C1-C5 hydrocarbons
Injection
Inj. Vol.:10 µL split
Liner:2 mm splitless (cat.# 20712)
Inj. Temp.:200 °C
Split Vent Flow Rate:35 mL/min
Oven
Oven Temp.:50 °C to 150 °C at 6 °C/min
Carrier GasHe, constant pressure (10 psi, 68.9 kPa)
Linear Velocity:47 cm/sec @ 50 °C
DetectorFID @ 200 °C
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentHP5890 GC