share:

This chromatogram may contain analytes that are synonyms of your search terms.

Refinery Gas

Peaks
1.Air
2.Methane
3.Carbon dioxide
4.Ethylene
5.Acetylene
6.Ethane
7.Propylene
8.Propane
9.Propadiene
10.Isobutane
11.Isobutylene
Peaks
12.1-Butene
13.1,3-Butadiene
14.n-Butane
15.cis-2-Butene
16.trans-2-Butene
17.Isopentane
18.1-Pentene
19.2-Methyl-2-butene
20.n-Pentane
21.cis-2-Pentene
22.n-Hexane
image
GC_PC01143
ColumnRt-QS-BOND, 30 m, 0.53 mm ID, 20 µm (cat.# 19738)
SampleRefinery gas standard
Injection
Inj. Vol.:20 µL split
Liner:2 mm (cat.# 20712)
Inj. Temp.:200 °C
Split Vent Flow Rate:35 mL/min
Oven
Oven Temp.:40 °C (hold 2 min) to 225 °C at 15 °C/min (hold 5 min)
Carrier GasHe, constant pressure (11.5 psi, 79.3 kPa)
Linear Velocity:68 cm/sec @ 40 °C
DetectorTCD @ 225 °C
Make-up Gas Type:He
Data Rate:20 Hz
Sensitivity Mode:He/H2
InstrumentHP5890 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Refniery Gas

,

Rt-QSBOND

,

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us