Gasoline Light Fractions on Rtx-DHA-150

Peaks
1.Methane
2.Ethane
3.Propylene
4.Propane
5.Isobutane
6.Butane
image
GC_PC1145
ColumnRtx-DHA-150, 150 m, 0.25 mm ID, 1.0 µm (cat.# 10149)
SampleGasoline
Conc.: Neat
Injection
Inj. Vol.:1.0 µL split (split ratio 150:1)
Liner:Cyclosplitter (cat.# 20706)
Inj. Temp.:275 °C
Oven
Oven Temp.:35 °C (hold 60 min) to 200 °C at 5 °C/min (hold 5 min)
Carrier GasH2, constant flow
Linear Velocity:44 cm/sec
DetectorFID @ 275 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
InstrumentAgilent/HP6890 GC
NotesLight fractions of gasoline above ambient temperature
Download PDF