Low Bleed, High Efficiency MXT-1HT SimDist Columns Outperform Competitors for Simulated Distillation (ASTM D6352)

image
GC_PC1149
ColumnRestek MXT-1HT SimDist (cat.#70115), see above for competitor column descriptions
SampleC8-C100 hydrocarbons
Diluent:CS2
Conc.:1.0%
Injection
Inj. Vol.:0.5 µL cold on-column
Temp. Program:53 °C to 430 °C at 10 °C/min (hold 5 min)
Oven
Oven Temp.:50 °C to 430 °C at 10 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:18 mL/min
DetectorFID @ 430 °C
Make-up Gas Flow Rate:24 mL/min
Make-up Gas Type:N2
InstrumentShimadzu 2010 GC