share:

This chromatogram may contain analytes that are synonyms of your search terms.

Hydrocarbons (C5-C106) on MXT-1HT SimDist at 430 °C

PeakstR (min)
1.----
2.----
3.----
4.----
5.----
6.----
7.----
8.C120.988
9.C131.677
10.C142.544
11.C153.503
12.C164.484
13.C175.451
14.C186.386
15.C208.15
16.C229.764
17.C2411.259
18.C2612.645
19.C2813.920
PeakstR (min)
20.C3015.134
21.C3216.280
22.C3417.364
23.C3618.392
24.C3819.371
25.C4020.363
26.C4221.199
27.C4422.041
28.C4622.843
29.C4823.609
30.C5024.340
31.C5225.041
32.C5425.712
33.C5626.365
34.C5826.990
35.C6027.594
36.C6228.173
37.C6428.733
38.C6629.274
39.C6829.797
PeakstR (min)
40.C7030.303
41.C7230.789
42.C7431.265
43.C7631.725
44.C7832.173
45.C8032.609
46.C8233.033
47.C8433.454
48.C8633.858
49.C8834.256
50.C9034.643
51.C9235.024
52.C9435.395
53.C9635.758
54.C9836.115
55.C10036.460
56.C10236.803
57.C10437.139
58.C10637.465
image
GC_PC1165
ColumnMXT-1HT SimDist, 5 m, 0.53 mm ID, 0.10 µm (cat.# 70112)
SampleCustom C5-C106 hydrocarbon standard
Diluent:Carbon disulfide
Conc.:1%
Injection
Inj. Vol.:0.5 µL cold on-column
Temp. Program:53 °C to 430 °C at 10 °C/min (hold 5 min)
Oven
Oven Temp.:50 °C to 430 °C at 10 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:18 mL/min
DetectorFID @ 430 °C
Make-up Gas Flow Rate:24 mL/min
Constant Column + Constant Make-up:42 mL/min
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentShimadzu 2010 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

crude Oil

,

SimDist

,

simulated distillation

,

crude

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us