share:

This chromatogram may contain analytes that are synonyms of your search terms.

Light Hydrocarbons on Rt-Alumina BOND/MAPD

Peaks
1.Isobutane
2.n-Butane
3.trans-2-Butene
4.1-Butene
5.Isobutylene
6.cis-2-Butene
7.1,2-Butadiene
8.n-Hexane
image
GC_PC1201
ColumnRt-Alumina BOND/MAPD, 50 m, 0.32 mm ID, 5.0 µm (cat.# 19780)
SampleCrude 1,2-butadiene
Diluent:None
Injection
Inj. Vol.:1 µL split
Liner:2.0 mm ID straight inlet liner (cat.# 20712)
Inj. Temp.:200 °C
Split Vent Flow Rate:80 mL/min
Oven
Oven Temp.:100 °C (hold 31 min)
Carrier GasH2, constant linear velocity
Linear Velocity:17.40 psi, 120.0 kPa @ 100 °C
DetectorFID @ 200 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
Data Rate:20 Hz
InstrumentHP5890 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

2-butadiene

,

MAPD

,

Alumina BOND

,

1

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us