Riser Pipe Oil from Deepwater Horizon Oil Spill on Rxi-17Sil and Rxi-1HT (GCxGC-TOFMS)

image
GC_PC1252
ColumnRxi-17Sil MS 60 m, 0.25 mm ID, 0.10 µm (cat.# custom)
Rxi-1HT 1 m, 0.25 mm ID, 0.10 µm (cat.# 13950)
SampleRiser pipe oil from Deepwater Horizon oil spill
SV internal standard mix (5 ng/µL) (cat.# 31206)
Diluent:Methylene chloride
Conc.:10 mg/mL
Injection
Inj. Vol.:1.0 µL split (split ratio 10:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.5)
Inj. Temp.:275 °C
Oven
Oven Temp.:Rxi-17Sil MS: 40 °C (hold 1 min) to 320 °C at 2.5 °C/min (hold 7 min)
Rxi-1HT: 45 °C (hold 1 min) to 325 °C at 2.5 °C/min (hold 7 min)
Carrier GasHe, corrected constant flow (1 mL/min)
Modulation
Modulator Temp. Offset:20 °C
Second Dimension Separation Time:2.8 sec
Hot Pulse Time:1.0 sec
Cool Time between Stages:0.4 sec
DetectorMS
Mode:
Transfer Line Temp.:300 °C
Analyzer Type:TOF
Source Temp.:250 °C
Electron Energy:-70 eV
Mass Defect:100 mu/100 u
Ionization Mode:EI
Acquisition Range:45 to 550 amu
Spectral Acquisition Rate:100 spectra/sec
InstrumentLECO Pegasus 4D GCxGC-TOFMS
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

pristane

,

phytane

,

iso-paraffins

,

olefins

,

cyclo-paraffins

,

alkyl-aromatics

,

alkyl-PAHs

,

dibenzothiophenes

,

steranes

,

hopanes

,

paraffins

,

aromatics

,

PAHs

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us