share:

Simulated Distillation on MXT-1HT SimDist by ASTM D7213 Using Nitrogen Carrier Gas

image
GC_PC1286
ColumnMXT-1HT SimDist, 5 m, 0.53 mm ID, 0.88 µm (cat.# 70131)
SampleRetention time standard for D7213 (C5-C72), Separation Systems, Inc.
Diluent:Carbon disulfide
Conc.:1 wt%
Injection
Inj. Vol.:1 µL PTV on column
Liner:Premium 1.7 mm ID on-column inlet liner (cat.# 23430.1)
Inlet Temp. Program:90 °C to 380 °C at 20 °C/min
Oven
Oven Temp.:40 °C to 380 °C at 10 °C/min (hold 10 min)
Carrier GasN2, constant flow
Flow Rate:11.9 mL/min
DetectorFID @ 390 °C
Make-up Gas Flow Rate:18 mL/min
Make-up Gas Type:N2
Hydrogen flow:43 mL/min
Air flow:400 mL/min
InstrumentAgilent 7890B GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Simulated Distillation

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.