share:

This chromatogram may contain analytes that are synonyms of your search terms.

Simulated Distillation on Rtx-1 SimDist 2887 DiatoSorb-WHP by ASTM D2887

Peaks
1.n-Pentane
2.n-Hexane
3.n-Heptane
4.n-Octane
5.n-Nonane
6.n-Decane
7.n-Undecane
8.n-Dodecane
9.n-Tetradecane
10.n-Pentadecane
Peaks
11.n-Hexadecane
12.n-Heptadecane
13.n-Octadecane
14.n-Eicosane
15.n-Tetracosane
16.n-Octacosane
17.n-Dotriacontane
18.n-Hexatriacontane
19.n-Tetracontane
20.n-Tetratetracontane
image
GC_PC1299
ColumnRtx-1 SimDist 2887, 100/120 mesh on DiatoSorb-WHP, SilcoSmooth tubing, 25 in, 1/8 in OD, 2.0 mm ID (cat.# 80000-800)
SampleASTM D2887-12 calibration standard (cat.# 31674)
Diluent:Carbon disulfide
Conc.:1 % w/w
Injection
Inj. Vol.:0.1 µL on-column
Inj. Temp.:350 °C
Oven
Oven Temp.:35 °C to 350 °C at 10 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:25 mL/min
DetectorFID @ 350 °C
InstrumentAgilent 7890B GC
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us