High-Temperature Simulated Distillation on MXT-1HT SimDist by ASTM D7169

image
GC_PC1308
ColumnMXT-1HT SimDist, 5 m, 0.53 mm ID, 0.10 µm (cat.# 70112)
SampleRetention time standard for D7169 (C5-C100)
Injection
Inj. Vol.:0.1 µL cool on-column
Temp. Program:50 °C to 425 °C at 15 °C/min
Oven
Oven Temp.:-20 °C to 425 °C at 15 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:20 mL/min
DetectorFID @ 435 °C
Make-up Gas Flow Rate:15 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
InstrumentAgilent 7890B GC