FAME in Soy B100 by EN 14103 (2011)

GC_PC1333
Column | FAMEWAX, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12497) |
Sample | Soy B100 biodiesel |
Methyl nonadecanoate (cat.# 35055) | |
Diluent: | Toluene |
Conc.: | 10 mg/mL, EN 14103 (2011) method preparation |
Injection | |
Inj. Vol.: | 1 µL split (split ratio 100:1) |
Liner: | Topaz 4.0 mm ID Precision inlet liner w/ wool (cat.# 23305) |
Inj. Temp.: | 240 °C |
Oven | |
Oven Temp.: | 60 °C (hold 2 min) to 200 °C at 10 °C/min to 240 °C at 5 °C/min (hold 7 min) |
Carrier Gas | H2, constant flow |
Flow Rate: | 1.7 mL/min |
Detector | FID @ 250 °C |
Instrument | Agilent 7890B GC |