share:

This chromatogram may contain analytes that are synonyms of your search terms.

FAME in Soy B100 by EN 14103 (2011)

PeakstR (min)Structural Nomenclature
1.Methyl palmitate15.149C16:0
2.Methyl stearate16.997C18:0
3.Methyl oleate17.171C18:1 (cis-9)
4.Methyl linoleate17.621C18:2 (cis-9,12)
5.Methyl nonadecanoate18.052C19:0
6.Methyl linolenate18.235C18:3 (cis-9,12,15)
image
GC_PC1333
ColumnFAMEWAX, 30 m, 0.25 mm ID, 0.25 µm (cat.# 12497)
SampleSoy B100 biodiesel
Methyl nonadecanoate (cat.# 35055)
Diluent:Toluene
Conc.:10 mg/mL, EN 14103 (2011) method preparation
Injection
Inj. Vol.:1 µL split (split ratio 100:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/ wool (cat.# 23305)
Inj. Temp.:240 °C
Oven
Oven Temp.:60 °C (hold 2 min) to 200 °C at 10 °C/min to 240 °C at 5 °C/min (hold 7 min)
Carrier GasH2, constant flow
Flow Rate:1.7 mL/min
DetectorFID @ 250 °C
InstrumentAgilent 7890B GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Biodiesel

,

EN14103

,

Fatty Acids Methyl Esters

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us