share:

This chromatogram may contain analytes that are synonyms of your search terms.

Permanent Gases on Rt-Msieve 5A (15 m x 0.25 mm x 20 μm)

PeakstR (min)Conc. (Mol%)
1.Helium0.58Balance
2.Argon0.883
3.Oxygen0.912
4.Nitrogen1.745
5.Methane2.344
6.Carbon monoxide7.625
image
GC_PC1349
ColumnRt-Msieve 5A, 15 m, 0.25 mm ID, 20 µm (cat.# 19773)
SamplePermanent gases (custom mix)
Injection
Inj. Vol.:30 µL split (split ratio 40:1)
Liner:Topaz 2.0 mm ID straight inlet liner (cat.# 23313)
Inj. Temp.:150 °C
Oven
Oven Temp.:30 °C (hold 10 min)
Carrier GasH2, constant pressure (8.0 psi, 55.2 kPa)
DetectorTCD @ 200 °C
InstrumentAgilent 7890B GC
NotesResolution (Ar/O2) = 1.7
Asymmetry (CO) = 1.2

Note that the sample was introduced onto the column using syringe injection and a high split ratio, which results in a very narrow sample band. Any dead volume in the system (e.g., from using a sampling valve, large sampling loops, or even from using splitless injection) will result in peak broadening, which might decrease resolution between the peaks or even result in peak coelution.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

MolSieve

,

Molecular Sieve

,

Natural Gas

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us