share:

This chromatogram may contain analytes that are synonyms of your search terms.

Glycerol and Glycerides in Biodiesel on Rtx-Biodiesel TG (10 m) by ASTM D6584

PeakstR (min)
1.Glycerol4.682
2.1,2,4-Butanetriol5.624
3.Monoglyceridesas shown
4.Tricaprin19.174
5.Diglycerides20.13 - 20.80
6.Triglycerides22.24 - 24.70
image
GC_PC1368
ColumnRtx-Biodiesel TG with 2 m x 0.53 mm ID retention gap, 10 m, 0.32 mm ID, 0.1 µm (cat.# 10291)
SampleSoy B100 biodiesel
Tricaprin (1,2,3-tricaprinoylglycerol) (cat.# 33025)
(s)-(-)-1,2,4-Butanetriol (cat.# 33024)
Diluent:Heptane
Injection
Inj. Vol.:1 µL cool on-column
Temp. Program:Track oven
Oven
Oven Temp.:50 °C (hold 1 min) to 180 °C at 15 °C/min to 230 °C at 7 °C/min to 380 °C at 30 °C/min (hold 10 min)
Carrier GasH2, constant flow
Flow Rate:1.6 mL/min
DetectorFID @ 380 °C
InstrumentAgilent 7890B GC
NotesSample preparation followed ASTM D6584-17: Soy biodiesel B100 was fortified with internal standards (butanetriol and tricaprin) and derivatized using MSTFA.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

B100

,

Bio Fuels

,

Monoglycerides

,

Diglycerides

,

Triglycerides

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us