share:

This chromatogram may contain analytes that are synonyms of your search terms.

Blood Alcohol on Rtx-BAC1 & Rtx-BAC2 (0.32mm ID Columns)

PeaksConc.
(wt.%)
1.Methanol0.1
2.Acetaldehyde0.2
3.Ethanol0.2
4.Isopropanol0.1
5.Acetone0.01
6.n-Propanol0.1
image
GC_PH00241
ColumnsRtx-BAC1 30 m, 0.32 mm ID, 1.80 µm (cat.# 18003)
and Rtx-BAC2 30 m, 0.32 mm ID, 1.20 µm (cat.# 18002)
using two hole ferrule
Sample1.0 mL headspace sample of a blood alcohol mix
Injection
Inj. Vol.:1,000 µL headspace-syringe split
Liner:2mm ID
Inj. Temp.:200 °C
Headspace-Syringe
Split Vent Flow Rate:20 mL/min
Sample Temp.:70 °C
Sample Equil. Time:15 min
Oven
Oven Temp.:40 °C
Carrier GasHe
DetectorFID
InstrumentPerkin Elmer HS 40 headspace autosampler
NotesVial pressure: 30 psi
Vial pressurization time: 0.15 min.
Vial sampling time: 0.01 min.
Transfer line: 0.32 mm ID Hydroguard® fused silica tubing
Transfer line temp.: 200°C
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

BAC

,

blood alcohol

,

DB-ALC1

,

DB-ALC2

,

alcohols

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us