Underivatized Ibuprofen on Rt-βDEXsm

image
GC_PH00242
ColumnRt-ßDEXsm, 30 m, 0.32 mm ID, 0.25 µm (cat.# 13104)
Sample
Conc.: 125 ng on-column each enantiomer
Injectionsplit (split ratio 13:1)
Liner:Cup Splitter (cat.# 20709)
Inj. Temp.:200 °C
Oven
Oven Temp.:175 °C to 200 °C at 2 °C/min
Carrier GasHe, constant flow
Linear Velocity:60 cm/sec
DetectorFID @ 230 °C
Download PDF