share:

Cocaine and Cocaine Adulterants Rtx-440

Peaks
1.Benzocaine
2.Phenacetin
3.Prilocaine
4.Lidocaine
5.Caffeine
6.Procaine
7.Cocaine
8.Tetracaine
image
GC_PH00769
ColumnRtx-440, 30 m, 0.25 mm ID, 0.50 µm (cat.# 12938)
Sample
Diluent:methanol
Conc.:100 µg/mL
Injection
Inj. Vol.:1.0 µL split (split ratio 10:1)
Liner:Laminar Cup Splitter (cat.# 20801)
Inj. Temp.:250 °C
Oven
Oven Temp.:150 °C to 275 °C at 25 °C/min to 300 °C at 15 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min
DetectorFID @ 300 °C
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

cocaine

,

coke

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.