share:

Basic Drugs (Underivatized) Rxi-5ms

Peaks
1.Amphetamine
2.Methamphetamine
3.Nicotine
4.Cotinine
5.Caffeine
6.Benzphetamine
Peaks
7.Ketamine
8.Phencyclidine
9.Methadone
10.Cocaine
11.Codeine
12.Scopolamine
13.Alprazolam
image
GC_PH00814
ColumnRxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423)
Sample1000μg/mL each
Diluent:methanol
Conc.:20 ng each compound on column
Injection
Inj. Vol.:1.0 µL split (split ratio 50:1)
Liner:Siltek Gooseneck Splitless (4mm) (cat.# 20799-214.5)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 220 °C at 15 °C/min to 330 °C at 10 °C/min (hold 5 min)
Carrier GasHe, constant pressure
Linear Velocity:30 cm/sec
DetectorAgilent 5973 MSD
Transfer Line Temp.:300 °C
Analyzer Type:Quadrupole
Tune Type:PFTBA
Ionization Mode:EI
Scan Range:35-550 amu
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

basic drugs

,

drugs of abuse

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.