Local Anesthetics Rxi-5ms

Peaks
1.Benzocaine
2.Prilocaine
3.Lidocaine
4.Procaine
5.Tetracaine
6.Bupivacaine
7.Dibucaine
image
GC_PH00860
ColumnRxi-5ms, 30 m, 0.53 mm ID, 1.00 µm (cat.# 13455)
Sample
Diluent:methanol
Conc.:50 µg/mL each component
Injection
Inj. Vol.:1.0 µL split (split ratio 10:1)
Liner:4mm Split w/Wool (cat.# 20781)
Inj. Temp.:250 °C
Oven
Oven Temp.:200 °C (hold 4 min) to 320 °C at 30 °C/min (hold 3 min)
Carrier GasHe, constant flow
Flow Rate:5.0 mL/min
DetectorFID @ 300 °C