share:

Steroid Sex Hormones on Rxi-1ms

Peaks
1.Androsterone
2.Dehydroepiandrosterone (DHEA)
3.17α-Estradiol
Peaks
4.Estrone
5.17β-Estradiol
6.Testosterone
7.derivatization by-product
image
GC_PH00872
ColumnRxi-1ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13323)
SampleEach hormone in methanol or ethanol; compounds derivatized using 2% methoxylamine HCl (CH3ONH2) in pyridine, then N-trimethylsilylimidazole (TMSI), then analyzed
Conc.:100 µg/mL
Injection
Inj. Vol.:1.0 µL splitless (hold 0.5 min)
Liner:3.5mm single gooseneck inlet liner
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 320 °C at 10 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min
DetectorMS-Scan: Shimadzu 17A with QP5000
Transfer Line Temp.:280 °C
Ionization Mode:EI
Scan Range:40-700 amu

RELATED CHROMATOGRAM LIBRARY SEARCHES

hormones

,

steroids

,

estradiol

,

TMSI

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us