share:

Over 10,000 Injections of Lipidomics on Rxi-5ms

Peaksm/z
1.Coprostanol370.40
2.d5-cholestanol (IS)220.20
3.7α-hydroxycholesterol456.40
4.Cholestanol306.30
5.d5-epicholestanol (IS)220.20
6.7-Dehydrocholesterol325.25
7.Desmosterol343.25
Peaksm/z
8.Lathosterol458.35
9.d4-lathosterol (IS)462.40
10.Campesterol382.35
11.4-Cholestenone382.30
12.Stigmasterol394.15
13.Lanosterol393.35
14.β-Sitosterol396.35
15.d7-β-sitosterol (IS)403.45
image
GC_PH01053
ColumnRxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423)
SampleLipid plasma extract as trimethylsilyl derivatives
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:4 mm split taper with wool
Inj. Temp.:310 °C
Oven
Oven Temp.:250 °C (hold 1 min) to 320 °C at 30 °C/min (hold 1.6 min)
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorMS
Mode:SIM
Transfer Line Temp.:330 °C
Ionization Mode:EI
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

lipidomics

,

trimethylsilyl derivatives

,

lipid plasma

,

sterol metabolites

,

cholesterol

,

TMS derivatives

,

Rxi-5ms

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us