share:

Mesylate, Besylate, and Tosylate Genotoxic Impurities on Rxi-624Sil MS (extracted ion chromatograms)

PeakstR (min)Ions Monitored (m/z)
1.Methyl methanesulfonate (MMS)5.04379, 80, 95
2.Isopropyl methanesulfonate (iPMS)6.01059, 79, 123
3.Diethyl sulfate (DES)6.357111, 125, 139
4.di-isopropyl sulfate (DPS)6.97045, 87, 167
5.dibutyl sulfate (DBS)9.26029, 41, 56
6.Methyl benzenesulfonate (MBS)9.46377, 141, 172
7.Ethyl benzenesulfonate (EBS)9.88377, 141, 186
8.Methyl toluenesulfonate (MTS)10.33791, 155, 186
9.Ethyl toluenesulfonate (ETS)10.71391, 155, 200
10.n-propyl toluenesulfonate (nPTS)10.82791, 155, 172
11.n-butyl benzenesulfonate (nBBS)11.0777, 141, 159
12.isopropyl p-toluenesulfonate (iPTS)11.24091, 155, 172
13.p-toluenesulfonic acid n-butyl ester (nBTS)11.82391, 155, 173
* ortho isomer of ETS
image
GC_PH1165
ColumnRxi-624Sil MS, 20 m, 0.18 mm ID, 1.00 µm (cat.# 13865)
Sample
Diluent:90:10 acetonitrile:water
Conc.:500 ng/mL
Injection
Inj. Vol.:1 µL splitless (hold 0.5 min)
Liner:3.5mm Single Gooseneck Liner with wool placed 3cm from top (middle) (cat.# 22286)
Inj. Temp.:220 °C
Purge Flow:3 mL/min
Oven
Oven Temp.:80 °C (hold 2 min) to 300 °C at 20 °C/min (hold 5 min)
Carrier GasHe, constant flow
Linear Velocity:45 cm/sec
DetectorMS
Mode:SIM
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Temp.:280 °C
Solvent Delay Time:4 min
Ionization Mode:EI
InstrumentShimadzu 2010 GC & QP2010+ MS
AcknowledgementIn collaboration with Merck and Company
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

GTI

,

mesylate

,

besylate

,

tosylate

,

PGI

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us