share:

Phenethyl Glucosinolate on Ultra Aqueous C18

Peaks
1.Phenethyl glucosinolate
image
LC_0166
ColumnUltra Aqueous C18 (cat.# 9178565)
Dimensions:150 mm x 4.6 mm ID
Particle Size:5 µm
Pore Size:100 Å
Temp.:ambient
Sample
Diluent:water
Conc.:1,000 µg/mL
Inj. Vol.:10 µL
Mobile Phase
A:50 mM potassium phosphate, pH 2.5
B:acetonitrile
Time (min)%B
0.000
1075
110
160
Flow:1.0 mL/min
DetectorUV/Vis @ 210 nm
Download PDF