Pesticides (Carbamates) on Ultra Carbamate (Fast LC)

Peaks
1.Aldicarb sulfone
2.Aldicarb sulfoxide
3.Oxamyl
4.Methomyl
5.3-Hydroxycarbofuran
Peaks
6.Aldicarb
7.Propoxur
8.Carbofuran
9.Carbaryl
10.Methiocarb
11.4-Bromo-3,5-dimethylcarbamate
image
LC_0225
ColumnUltra Carbamate (cat.# 9177355)
Dimensions:50 mm x 4.6 mm ID
Particle Size:3 µm
Pore Size:100 Å
Temp.:27 °C
Sample4-Bromo-3,5-dimethylphenyl-N-methylcarbamate (BDMC) (cat.# 32274)
531.1 Carbamate pesticide calibration mixture (cat.# 32273)
Diluent:Methanol
Conc.:50 µg/mL each
Inj. Vol.:5 µL
Mobile Phase
A:90:10 Water:methanol
B:90:10 Methanol:acetonitrile
Time (min)%B
0.0010
1090
Flow:1.5 mL/min
DetectorUV/Vis @ 220 nm
NotesStandards mixed 50:50