Pesticides (Carbamates) on Ultra Carbamate

PeaksIonCone V
1.Aldicarb sulfone223.325V
2.Aldicarb sulfoxide207.318V
3.Oxamyl237.2*10V
4.Methomyl163.215V
5.3-Hydroxycarbofuran238.315V
6.Aldicarb191.28V
7.Propoxur210.218V
8.Carbofuran222.322V
9.Carbaryl202.218V
10.Methiocarb226.319V
*Ammonium adduct (all other are [M+H]+ ions)
image
LC_0308
ColumnUltra Carbamate (cat.# 9177315)
Dimensions:100 mm x 4.6 mm ID
Particle Size:3 µm
Pore Size:100 Å
Sample
Conc.:100 µg/mL each component
Inj. Vol.:10 µL
Mobile Phase
A:90% water:10% methanol with 10mM ammonium formate
B:90% acetonitrile:10% methanol with 10mM ammonium formate
Time (min)%B
010
1590
2090
Flow:1.0 mL/min
DetectorUV/Vis @ 220 nm
NotesFlow: (approx. 0.75mL/min. to UV detector, 0.25mL/min. to MSD)

MSD Conditions:
Detector: Micromass ZMD
Mode: ESI+
Capillary V: 3.50
Extractor: 4.0
Ion energy: 0.4
Multiplier: 650
Source temp: 100°C
Desolv. temp: 250°C
Gas flow: 490 L/hr.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Carbamates

,

Ultra Carbamate

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us