share:

100 ng/mL Benzodiazepines in Hydrolyzed Urine on Ultra Biphenyl by LC-MS/MS

PeakstR (min)MRM1MRM2MRM3
1.7-Aminoclonazepam0.59286.0/250.3286.0/222.2286.0/121.0
2.7-Aminoflunitrazepam0.77284.2/135.2284.2/240.2284.2/226.1
3.Bromazepam1.10315.9/182.0315.9/209.1315.9/260.9
4.N-Desmethylflunitrazepam1.71300.0/254.1300.0/225.2300.0/198.0
5.Nitrazepam1.69282.1/236.1282.1/207.2282.1/180.0
6.Lorazepam1.71321.0/302.9321.0/275.1321.0/229.1
7.2-Hydroxyethylflurazepam1.79333.1/246.1333.1/166.0333.1/109.1
8.Nordiazepam1.78271.0/140.0271.0/208.0271.0/225.9
9.Estazolam1.88294.9/267.1294.9/241.0294.9/138.0
10.Desalkylflurazepam1.90289.0/140.0289.0/165.1289.0/226.0
11.Clonazepam1.92315.9/270.2315.9/214.1315.9/207.1
12.Alprazolam1.99309.4/281.1309.4/205.0309.4/274.1
13.Midazolam1.26326.1/291.2326.1/244.2326.1/209.0
14.Triazolam2.06343.1/315.0343.1/308.2343.1/239.2
15.Temazepam2.13301.2/283.1301.2/255.1301.2/228.1
16.Flurazepam1.34388.1/315.1388.1/288.2388.1/317.1
17.Flunitrazepam2.24314.4/211.0314.4/239.1314.4/268.1
18.Clobazam2.26301.1/259.1301.1/224.0301.1/181.1
19.Diazepam-D5 (IS)2.42290.0/154.1----
20.Diazepam2.45284.8/193.1284.8/222.2284.8/257.1
image
LC_CF0525
ColumnUltra Biphenyl (cat.# 9109552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:100 Å
Temp.:40 °C
Sample
Diluent:Starting mobile phase + 30 ng/mL IS
Conc.:100 ng/mL sample was diluted 20x prior to injection
Inj. Vol.:30 µL
Mobile Phase
A:Water + 2 mM ammonium formate + 0.2% formic acid
B:Acetonitrile + 10% water + 2 mM ammonium formate + 0.2% formic acid
Time (min)Flow (mL/min)%A%B
0.000.67030
3.00.64060
3.10.6595
3.50.6595
3.60.67030
4.5stop
DetectorAB SCIEX API 4000 MS/MS
Model #:API 4000
Ion Source:TurboIonSpray®
Ion Mode:ESI+
Ion Spray Voltage:2 kV
Curtain Gas:40 psi (275.8 kPa)
Gas 1:40 psi (275.8 kPa)
Gas 2:40 psi (275.8 kPa)
Source Temp.:600 °C
Mode:MRM
InstrumentApplied Biosystems/MDS Sciex LC-MS/MS System
NotesMS/MS instrument was operated in scheduled MRM mode.
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us