share:

This chromatogram may contain analytes that are synonyms of your search terms.

Norbuprenorphine & Buprenorphine in Urine on Raptor Biphenyl by LC-MS/MS

PeakstR (min)Parent IonProduct Ion 1Product Ion 2
1.Norbuprenorphine0.63414.183.4101.0
2.Buprenorphine0.84468.3396.4414.5
image
LC_CF0580
ColumnRaptor Biphenyl (cat.# 9309A5E)
Dimensions:50 mm x 3.0 mm ID
Particle Size:2.7 µm
Temp.:30 °C
Sample
Conc.: 100 ng/mL standard was prepared in 1 mL urine. The 1 mL urine standard was then diluted with 2 mL of 0.1% formic acid in methanol:water (50:50) prior to analysis.
Inj. Vol.:10 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in methanol
Time (min)Flow (mL/min)%A%B
0.000.754555
1.00.750100
1.50.750100
1.510.754555
2.00.754555
DetectorAB SCIEX API 4000 MS/MS
Ion Source:TurboIonSpray®
Ion Mode:ESI+
InstrumentAPI LC-MS/MS
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Norbuprenorphine

,

buprenorphine

,

,

Raptor

,

Biphenyl

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us