share:

THC + Metabolites in Urine on Raptor Biphenyl by LC-MS/MS

PeakstR (min)Q1Q3 QuantifierQ3 Qualifier
1.THC-OH-d3 (IS)1.94334.35196.26-
2.THC-OH1.94331.32193.18105.11
3.THC-COOH-d3 (IS)2.09348.35196.24-
4.THC-COOH2.09345.36193.18119.16
5.Δ9-THC-d3 (IS)2.32318.42196.25-
6.Δ9-THC2.33315.32193.18135.16
image
LC_CF0583
ColumnRaptor Biphenyl (cat.# 9309A5E)
Dimensions:50 mm x 3.0 mm ID
Particle Size:2.7 µm
Guard Column:Raptor Biphenyl EXP 5 mm, 3.0 mm ID, 2.7 µm (cat.# 9309A0253)
Temp.:40 °C
SampleCustom
Conc.: A 50 ng/mL standard was prepared in urine. 1 mL of the 50 ng/mL urine standard was diluted with 2 mL of 0.1% formic acid in 50:50 water:methanol prior to analysis.
Inj. Vol.:2 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in methanol
Time (min)Flow (mL/min)%A%B
0.000.7004060
2.50.7000100
3.00.7000100
3.010.7004060
4.00.7004060
DetectorWaters Xevo TQ-S
Ion Source:Waters Zspray™ ESI
Ion Mode:ESI+
InstrumentWaters
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

THC

,

metabolites

,

Raptor

,

,

LC-MS/MS

,

urine

,

cannabinoids

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us