share:

This chromatogram may contain analytes that are synonyms of your search terms.

Drugs of Abuse on Raptor Biphenyl by LC-MS/MS

PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
1.Morphine0.66286.2152.3165.3
2.Amphetamine0.78136.0119.191.0
3.Methamphetamine0.93150.091.2119.3
4.Codeine1.00300.2165.2152.2
5.6-MAM1.01328.1165.1211.1
6.MDMA1.06194.1163.1105.1
7.Benzoylecgonine1.33290.1168.1105.1
8.Cocaine1.39304.2182.182.1
9.Cocaethylene1.49318.2196.282.2
10.PCP1.64244.286.1159.2
image
LC_CF0587
ColumnRaptor Biphenyl (cat.# 9309552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:90 Å
Temp.:30 °C
Sample
Diluent:Water
Conc.:5 ng/mL
Inj. Vol.:10 µL
Mobile Phase
A:Water + 0.1% formic acid
B:Methanol + 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.89010
2.000.80100
2.010.89010
3.500.89010
DetectorMS/MS
Ion Mode:ESI+
InstrumentHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

illegal drugs

,

illegal

,

opiates

,

amphetamines

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.