share:

This chromatogram may contain analytes that are synonyms of your search terms.

Naltrexone and 6-β-Naltrexol in Urine on Raptor Biphenyl by LC-MS/MS

PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
1.Naltrexone0.90342.4270.1267.1
2.6-β-Naltrexol0.98344.4308.0254.1
image
LC_CF0588
ColumnRaptor Biphenyl (cat.# 9309552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:90 Å
Temp.:30 °C
Sample
Conc.: Prepared 500 ng/mL in human urine. Diluted 5x in mobile phase A. Final conc. = 100 ng/mL
Inj. Vol.:10 µL
Mobile Phase
A:Water + 0.1% formic acid + 2 mM ammonium formate
B:Water:methanol (5:95) + 0.1% formic acid + 2 mM ammonium formate
Time (min)Flow (mL/min)%A%B
0.000.57030
1.500.50100
1.510.57030
3.000.57030
Max Pressure:110 bar
DetectorMS/MS
Ion Mode:ESI+
InstrumentHPLC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

alcohol addiction

,

alcohol

,

alcohol dependence

,

treatment

,

alcohol treatment

,

opioid

,

opioid dependence

,

opioid treatment

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us