share:

This chromatogram may contain analytes that are synonyms of your search terms.

Stimulant Drug Panel on Raptor Biphenyl by LC-MS/MS

PeakstR (min)Precursor IonProduct IonProduct Ion
1.Nicotine1.35163.2117.1130.1
2.Norcotinine1.36163.180.1118.1
3.trans-3-Hydroxycotinine1.41193.180.0134.0
4.Anabasine1.50163.2120.180.1
5.Lisdexamfetamine1.58264.284.191.0
6.BZP1.64177.291.165.1
7.1S,2R-(+)-Ephedrine1.88166.2148.1115.0
8.Amphetamine2.03136.291.065.1
9.DMAA2.06116.157.199.2
10.Methylephedrine2.08180.2162.0115.1
11.Cotinine2.15177.280.098.1
12.Methamphetamine2.24150.391.1119.0
13.MDA2.28180.1163.1105.2
14.Phentermine2.33150.291.1133.1
15.Methylone2.33208.1160.1132.1
16.MDMA2.43194.2163.0135.1
17.Levamisole2.47205.1178.091.1
18.Diethylpropion2.50206.2105.1100.1
19.m-Hydroxybenzoylecgonine2.52306.1168.1121.0
20.Mephedrone2.52178.2160.1145.0
21.Ritalinic acid2.60220.184.156.2
22.MDEA2.60208.1163.0105.2
23.mCPP2.72197.1154.0118.1
24.Benzoylecgonine2.83290.1168.177.1
25.Methylphenidate2.85234.184.156.1
26.Cocaine2.96304.2182.077.1
27.Norcocaine3.02290.1168.0136.1
28.MDPV3.10276.2126.1135.1
29.Cocaethylene3.15318.2196.182.0
image
LC_CF0617
ColumnRaptor Biphenyl (cat.# 9309A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Guard Column:Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252)
Temp.:30 °C
Sample
Diluent:Water
Conc.:25 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water + 0.1% formic acid + 5 mM ammonium formate
B:Methanol + 0.1% formic acid + 5 mM ammonium formate
Time (min)Flow (mL/min)%A%B
0.000.6955
3.500.61090
3.510.6955
5.500.6955
DetectorMS/MS
Ion Mode:ESI+
Mode:Scheduled MRM
InstrumentUHPLC
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us