share:

This chromatogram may contain analytes that are synonyms of your search terms.

The Analysis of Morphine & Metabolites in Urine without Hydrolysis on Force Biphenyl

PeakstR (min)Precursor IonProduct IonProduct Ion
1.Morphine-3-β-D-glucuronide0.61462.1286.0152.1
2.Morphine1.06286.1152.0164.9
3.Morphine-6-β-D-glucuronide1.17462.1286.0152.1
4.Morphine-N-oxide1.30302.1284.9161.9
5.6-Monoacetylmorphine1.81328.1211.1165.0
6.Hydrocodone1.91300.3199.2128.3
image
LC_CF0680
ColumnForce Biphenyl (cat.# 9629252)
Dimensions:50 mm x 2.1 mm ID
Particle Size:1.8 µm
Guard Column:UltraShield UHPLC precolumn filter, 0.2 µm frit (cat.# 25809)
Temp.:40 °C
Sample
Conc.: A 750 ng/mL standard was prepared in 1 mL of urine. 100 μL of the fortified urine standard was diluted with 900 μL of 0.1% formic acid in water prior to analysis.
Inj. Vol.:2 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in methanol
Time (min)Flow (mL/min)%A%B
0.000.58515
0.500.58515
2.000.53070
2.010.58515
4.000.58515
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us