share:

This chromatogram may contain analytes that are synonyms of your search terms.

Metanephrine, Normetanephrine, and 3-Methoxytyramine in Human Plasma on Raptor HILIC-Si

PeakstR (min)Conc. (pg/mL)Precursor IonProduct Ion
1.3-Methoxytyramine-d4 (IS)1.80400155.07122.93
2.3-Methoxytyramine1.8050151.00119.00
3.Metanephrine-d3 (IS)2.03200183.00151.15
4.Metanephrine2.0350179.94148.22
5.Normetanephrine-d3 (IS)2.11400169.00136.96
6.Normetanephrine2.1150166.00134.02
image
LC_CF0689
ColumnRaptor HILIC-Si (cat.# 9310A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:30 °C
Sample
Diluent:Mobile phase A:mobile phase B (10:90)
Inj. Vol.:10 µL
Mobile Phase
A:Water, 100 mM ammonium formate, pH 3.0
B:Acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.31090
5.000.31090
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
NotesSolid Phase Extraction Procedure:
Charcoal stripped human plasma was fortified with analytes at 50 pg/mL (0.25, 0.27, and 0.30 nmol/L for metanephrine, normetanephrine, and 3-methoxytyramine, respectively). Internal standard (IS) was prepared at 4 ng/mL for metanephrine-d3 and at 8 ng/mL for normetanephrine-d3 and 3-methoxytyramine-d4 in methanol. The plasma sample (200 μL) was mixed with 10 μL of IS solution and 600 μL of 50 mM ammonium acetate solution. The mixture was loaded in an EVOLUTE EXPRESS WCX 96-well plate (30 mg) and washed with 1 mL water and 1 mL methanol:acetonitrile (50:50). The elution was performed twice with 0.9 mL of 5% formic acid in methanol:acetonitrile (50:50) and evaporated to dryness at 55 °C under a gentle stream of nitrogen. Dried extract was reconstituted with 100 μL of diluent and injected (10 μL) for analysis.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

neurotransmitters

,

monoamine

,

metanephrine

,

normetanephrine

,

3-methoxytyramine

,

HILIC

,

catecholamine

,

Raptor

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.