share:

Creatine and Creatinine in Human Urine on Raptor HILIC-Si by LC-MS/MS

PeakstR (min)Precursor IonProduct IonProduct Ion
1.Creatinine0.537114.044.386.0
2.Creatine1.240132.143.390.2
image
LC_CF0723
ColumnRaptor HILIC-Si (cat.# 9310A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:UltraShield UHPLC precolumn filter 0.2 µm (cat.# 25810)
Temp.:40 °C
Sample
Diluent:20:80 Water:acetonitrile
Conc.: Endogenous levels
Inj. Vol.:0.2 µL
Mobile Phase
A:5 mM Ammonium formate in 20:80 water:acetonitrile
Time (min)Flow (mL/min)%A
0.000.5100
1.5Stop
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
NotesSample Preparation:
Endogenous levels of creatinine and creatine in human urine were determined using a simple dilute-and-shoot method. A 50 μL aliquot of human urine was mixed with 950 μL acetonitrile. After vortexing and centrifuging at 4,300 rpm for 10 min, 10 μL of the supernatant was added to the 1,490 μL of 80% acetonitrile in water. Centrifugation was performed again before injection.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Raptor

,

HILIC-Si

,

Creatine

,

Creatinine

,

human

,

urine

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.